High-resolution silicon photonics focused ultrasound transducer with a sub-millimeter aperture
Optics Letters
         
          10
        
          48
        
          2668-2671
         
          2023
      
    
  
    Type: Zeitschriftenaufsatz (reviewed)
  
  
          DOI: 10.1364/OL.486567
      
  
    Abstract    
      
          We present an all-optical focused ultrasound transducer with a sub-millimeter aperture and demonstrate its capability for high-resolution imaging of tissue ex vivo. The transducer is composed of a wideband silicon photonics ultrasound detector and a miniature acoustic lens coated with a thin optically absorbing metallic layer used to produce laser-generated ultrasound. The demonstrated device achieves axial resolution and lateral resolutions of 12 μm and 60 μm, respectively, well below typical values achieved by conventional piezoelectric intravascular ultrasound. The size and resolution of the developed transducer may enable its use for intravascular imaging of thin fibrous cap atheroma.